Induction Heater | Working Voltage: 220V, 50/ 60Hz, Three Phase Working Current: 5~52 Amps (60A breaker required) Max. Input Power: 15 KW Output Frequency: 30 - 80 kHz Duty Cycle: 80% Induction Coil: 150mm(OD) x 140mm(ID) x 65mm(H) Air Cooling: Rear Fan Protection: Automatic water pressure, over-temperature, and overpower protection Refractory Assembling: Complete refractory assembling made of alumina, including bottom crucible support, crucible holder and cover
|
Temp. Controller & Thermocouple | Precision digital temperature controller with Automatic PID tunning 30 segments programmable for controlling heating and cooling rate as well as dwell time. Overheating & thermocouple failure protection. +/- 3℃ temperature accuracy. Thermocouple: B type (1/4"x10"L) Continuous Working Temp. Range: 1000℃~2000℃ Max. Heating Rate:
|
Graphite Crucible & Refractory liners | High Purity Graphite Crucible: 2" OD x 1.57"ID x 4" Depth (4.3'' Height) Crucible with multi-cavity is included for induction metal melting. 6 mm dia. Cavity: 16 8 mm dia. Cavity: 12
Three-layer special design refractory liner can withstand temperature up to 2000℃.
|
Sealing Flanges & Quartz Chamber | Two-end Open Quartz Tube is included with dimensions of 130mm(OD) x 122mm(ID) x 360mm(L) Top Flange: Stainless steel hinged type flange with 1/4'' thermocouple feedthrough, needle valve and bard fitting. Bottom Flange: Stainless steel flange with needle valve and KF-25 vacuum port. KF25 right angle valve& clamps and 1000mm long SS bellows are included for connecting to vacuum pum
|
Pressure Sensor | Anti-Corrosive Capacitance Diaphragm Gauge (Gas Independent), measurement range upto 3.8x10E-5 torr One 24V,1A DC power adapter is included and ready to use. Input voltage for the adapter is 110~240V Single Phase.
|
Flow Meter
| |
Water Chiller | Temperature range: 5 - 30℃ Water Flow Rate: > 16 L/m, Water pressure: 40PSI Note: It is recommended to use specific coolant to get the best performance.
|
Vacuum Pump | One 156L/m Rotary Vane Vacuum Pump is included Rate: 156 L/min up to 10e-2 torr vacuum. Max. Power Consumption: 0.4KW / 0.37KW For US customer, 110V Vacuum pump will be included If you choose high-speed turbo-pump, vacuum level can reach 10- 6 torr
|
Heavy Duty Mobile Cart | |
Protective Shield | For Non-UL Certified version, a protective shield made of phenolic sheets will be provided. For UL Certified version, a clear protective shield made of polycarbonate will be provided. The protective shield is to be placed around the quartz tube for safe operation
|
Product Dimensions | Induction Heater: 18.5"(L) x 12"(W) x 17"(H) (520mm x 520mm x 450mm) Water Chiller: 18"(L) x 18.7"(W) x 30"(H) (460mm x 475mm x 762mm) Heavy Duty Mobile Cart: 600mm(L) x 800mm(W) x 597mm(H)
|
Warranty | One year limited warranty with lifetime support Consumable parts such as processing tube, refractory liners, o-rings, and heating coil are not covered by the warranty
|
Warning | Attention: Protective inert gas must be continuously filled throughout the experiment in order to protect the graphite crucible and C thermocouple. The flow rate for gases should be limited to < 200 SCCM ( or 200ml/min) The Induction melting system is designed for use under vacuum and low pressure < 0.02 MPa Attention: A two-stage pressure regulator must be installed on gas cylinder to limit the pressure to below 3 PSI for safe operation. Never heat the furnace while the gas release valve is closed and a positive pressure condition exists within the furnace chamber. The pressure gauge must be used to closely monitor the chamber condition at all times during the heat process
|
Net Weight | |